Description
Available with chamber size of either 17 litre or 22 litre.
Supplied with reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test.
W&H service plans are also available – Premium Care and Premium Careplus.
-
Lyla sterilization programs
Sterilization cycle S Wrapped 134 S Prion 134 S Gentle 121 S Unwrapped 134 Temperature: 134 °C 134 °C 121 °C 134 °C Holding time (minutes): 4 18′ 30” 20′ 30” 4 Total cycle duration
(minutes) including
complete drying:Automatic from Typical
loadAutomatic from Typical
loadAutomatic from Automatic from Typical
loadLyla 17 38′ to 45′42′ 52′ to 59′56′ 61′ to 64′ 20′ to 26′ 23′ Lyla 22 (220 V) 39′ to 46′43′ 53′ to 60′57′ 62′ to 66′ 21′ to 27′ 24′ Lyla 22 (110 V) 49′ to 56′51′ 63′ to 71′65′ 69′ to 73 28′ to 35′ 30′ Lyla 17/22 loads:maximum solid 4 kg/typical 2 kg/maximum porous 0.5 kgmaximum 4 kg unwrappedLyla 17/22 container load: maximum 4 kg not applicable Test cycles: Bowie and Dick/Vacuum test Item supplied with reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test
Lyla Sterilizer
Lyla 17 Lyla 22 Type: RIN-210 RIN-210 Chamber size: 17 l 22 l Power supply: 200–240 V AC; 50 Hz or 60 Hz; 10 A single phase 200–240 V AC; 50 Hz or 60 Hz; 10 A single phase
100–125 V AC; 50 Hz/60 Hz; 12 A single phasePower consumption: 2.0–2.4 kW 220 V: 2.0–2.4 kW
110 V: 1.2–1.5 kWOverall dimension
(width × height × depth):465 × 452 × 646 mm Weight: 40 kg 42 kg Main / used water tanks: 4.8/4.8 l Working range: 7 to 10 cycles Usable space in chamber:
(width × height × depth)¹:195 × 195 × 312 mm 195 × 195 × 400 mm Connection types: 1 USB port in the front (additional 1 in the rear as option), automatic water filling kit (optional) (1) Usable space is referred to standard rack configuration.
Lisa sterilizers were designed, certified and validated with the most stringent directives and standards: N°2017/745 Medical Device Regulation 2014/68/EU Pressure Equipment Directive EN 13060 Small steam sterilizer IEC 61326-1 Electromagnetic compatibility IEC 61010-1 Safety requirements sterilizer IEC 61010-2-040 Specific requirements for steam sterilizer IEC 61770 Electric appliances connected to the water mains The sterilizer can be validated according to EN ISO 17665-1