Description
Available with chamber size of either 17 litre or 22 litre.
Supplied with reversible rack, 5 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test.
24 months’ warranty.
W&H service plans are also available – Premium Care and Premium Careplus.
Lisa sterilization cycles
Sterilization cycle | B-Universal 134 | B-Prion 134 | B-Universal 121 | Fast cycle | |||
---|---|---|---|---|---|---|---|
Temperature: | 134 °C | 134 °C | 121 °C | 134 °C | |||
Holding time (minutes): | 4 | 18.5 | 20.5 | 3.5 | |||
Total cycle duration (minutes) including complete drying: |
![]() |
Typical load |
![]() |
Typical load |
![]() |
![]() |
|
Lisa 17 |
Automatic from
21 to 41
|
28 |
Automatic from
36 to 57
|
43 | 35 to 67 | 13 to 20 | |
Lisa 22 |
Automatic from
21 to 45
|
28 |
Automatic from
36 to 61
|
43 | 40 to 72 | 13 to 21 | |
Lisa 17 loads:
Lisa 22 loads:
|
maximum solid 4.5 kg/typical 2 kg/maximum porous 1.5 kg
maximum solid 6 kg/typical 2 kg/maximum porous 2 kg
|
maximum solid 2 kg
maximum solid 2 kg
|
|||||
Lisa 17 / 22 containers load: | maximum 9 kg | – | |||||
Test cycles: | Helix/Bowie Dick/Vacuum test |
Items supplied with reversible rack, 5 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test
Lisa Sterilizer
Lisa 17 | Lisa 22 | |
---|---|---|
Type: | VA131-17 | VA131-22 |
Chamber size: | 17 l | 22 l |
Power supply: | 200–240 V AC; 50/60 Hz; 10 A | |
Power consumption: | 2.0 – 2.4 kW | |
Overall dimension (width x height x depth): |
465 x 452 x 634 mm | |
Weight: | 46 kg | 47.5 kg |
Main / Used water tanks: | 4.8/4.8 l | |
Working range: | 7 to 15 cycles | |
Usable space in chamber (width x height x depth)¹: |
195 x 195 x 312 mm | 195 x 195 x 400 mm |
Connection types: | 1 Ethernet port, 5 USB ports, integrated automatic water filling valve |
(1) Usable space is referred to standard rack configuration.
93/42/CEE | Medical Device Directive |
2014/68/EU | Pressure Equipment Directive |
EN 13060 | Small steam sterilizer |
IEC 61326-1 | Electromagnetic compatibility |
IEC 61010-1 | Safety requirements |
IEC 61010-2-040 | Specific requirements for steam sterilizer |
IEC 61770 | Electric appliances connected to the water mains |
The sterilizer can be validated according to EN ISO 17665-1